GuidesClasses
Cleanroom classes: ISO 1 to ISO 9
A class is the most particles allowed in one cubic metre of air, at set sizes. ISO 1 is the cleanest and ISO 9 the least clean. Each class up allows ten times more particles.

The limits
Particles per cubic metre, at or above each size. A dot means the standard sets no limit there.
| Class | 0.1 µm | 0.2 µm | 0.3 µm | 0.5 µm | 1 µm | 5 µm |
|---|---|---|---|---|---|---|
| ISO 1 | 10 | 2 | ||||
| ISO 2 | 100 | 24 | 10 | 4 | ||
| ISO 3 | 1,000 | 237 | 102 | 35 | 8 | |
| ISO 4 | 10,000 | 2,370 | 1,020 | 352 | 83 | |
| ISO 5 | 100,000 | 23,700 | 10,200 | 3,520 | 832 | |
| ISO 6 | 1,000,000 | 237,000 | 102,000 | 35,200 | 8,320 | 293 |
| ISO 7 | 352,000 | 83,200 | 2,930 | |||
| ISO 8 | 3,520,000 | 832,000 | 29,300 | |||
| ISO 9 | 35,200,000 | 8,320,000 | 293,000 |
Source: ISO 14644-1:2015 Table 1. The limit for class N at size D µm is 10N × (0.1 / D)2.08, rounded to three figures.
Old names: Class 100 to Class 100,000
The US standard FED-STD-209E counted particles of 0.5 µm and larger per cubic foot. It was cancelled in 2001, but the names stuck.
| Old name | ISO class |
|---|---|
| Class 1 | ISO 3 |
| Class 10 | ISO 4 |
| Class 100 | ISO 5 |
| Class 1,000 | ISO 6 |
| Class 10,000 | ISO 7 |
| Class 100,000 | ISO 8 |
GMP grades A to D
Medicine makers use grades, not classes. Each grade has a limit at rest and a looser one in operation.
| Grade | At rest | In operation | Used for |
|---|---|---|---|
| Grade A | ISO 5 | ISO 5 | Open sterile product: filling, stoppering, aseptic connections |
| Grade B | ISO 5 | ISO 7 | The room around a Grade A zone |
| Grade C | ISO 7 | ISO 8 | Less critical steps, such as preparing solutions to be filtered |
| Grade D | ISO 8 | Set by the maker | Handling parts after washing |
Source: EU GMP Annex 1 (2022), which PIC/S adopted. Malaysia is a PIC/S member; NPRA inspects to it.
What each class is used for
| Class | Often used for |
|---|---|
| ISO 4 | Wafer lithography; minienvironments around wafers |
| ISO 5 | The critical zone: sterile filling, laminar flow benches, wafer handling |
| ISO 6 | Chip assembly, optics |
| ISO 7 | Medical device assembly, pharmacy compounding, electronics |
| ISO 8 | Gowning rooms, packaging, electronics assembly |
Typical uses, not rules. Your process, customer or regulator sets the class.
Questions
What is a 10K cleanroom?
A Class 10,000 room under the old US standard: no more than 10,000 particles of 0.5 µm and larger per cubic foot. Today that is ISO 7.
Is ISO 8 the same as Class 100,000?
Close enough to use as one. ISO 8 allows 3,520,000 particles of 0.5 µm and larger per cubic metre. Class 100,000 allowed 100,000 per cubic foot, about 3,530,000 per cubic metre.
Which is cleaner, ISO 7 or ISO 8?
ISO 7. It allows ten times fewer particles than ISO 8: 352,000 against 3,520,000 per cubic metre at 0.5 µm.
What is the cleanest cleanroom class?
ISO 1. Few rooms run it. Wafer fabs use ISO 3 to ISO 5 around the wafer, often inside a minienvironment rather than the whole room.